CE Approval , Portable Medical Ventilator AV-2000B1
  • CE Approval , Portable Medical Ventilator AV-2000B1

Medical Ventilator suitable for pediatric and adult patient. be widely used for first aid, ambulance, Emergency situation ,ICU treatment,patients transport in hospital. Be equiped with Oxygen Cylinder, ventilator. oxygen Bridge

Features Specification Configuration Product Image Applications

Portable Medical Ventilator  AV-2000B1

Medical Ventilator suitable for pediatric  and adult patient.  be widely used for first aid, ambulance, Emergency situation ,ICU treatment,patients transport in hospital.

Be equiped with Oxygen Cylinder, ventilator. oxygen Bridge

Certification: CE /ISO13485

 

Specification:

1.Display mode              High-definition 5.7” LCD screen display

2.Ventilation mode       IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH

3.Ventilation parameters           

Tidal volume         50~1500ml

Rate                 2~99bpm

SIMV rate            2~20bpm

I:E                  2:1~1:8

Inspiratory trigger pressure -1.0~2.0 kPa

PEEP                0~2.0 kPa

Pressure range        1.0~6.0 kPa

SIGH          1.5 times the inspiratory time

 

Parameters for ventilation monitoring  tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, (oxygen concentration)

 

4. Security Alarm System

Oxygen Concentration alarm     

upper limit setting range  (50%~100%)

low limit setting range    (15%~50%)

Airway pressure alarm           

upper limit setting range       1.0~6.0 kPa

low limit setting range        0.4~2.0 kPa

Per-minute ventilation volume alarm   

upper limit setting range       3.0~30L/min

low limit setting range        1.0~10L/min

Sustained high-pressure alarm    

 it will give alarm when stress have consistently been higher than 2.5 kPa

Suffocation alarm              

it will give sound and light alarm if there is no tidal volume input for 15 seconds

Power alarm

 

5. Power        AC 220V±10%     50Hz±1Hz

                               

UPS, Storage battery

*Items within brackets are optional.